|
Frequently Asked Questions AboutFluorometric
Chlorophyll Analysis
Q: Why measure chlorophyll?
A: All plant life contains the primary photosynthetic pigment
chlorophyll a. Microscopic, planktonic plants, or
phytoplankton, occupy the lit zone of all water bodies. With over 70%
of the surface of the earth covered in water, phytoplankton and
photosynthetic bacteria are responsible for almost ½ of the
planets primary production while their total biomass comprises less
then 1% of the total plant biomass. These extraordinarily efficient
plants also act as the single largest CO2 sink
on earth. For these reasons alone it should be clear that there is an
interest in measuring concentrations of phytoplankton. Chlorophyll a
fluorescence is the most versatile, sensitive and easy
way to measure the concentrations of phytoplankton in water.
The quantitation, through extracted analysis, or
estimation, through in vivo analysis, of
chlorophyll a concentration supplies information
on the abundance of phytoplankton present in all aquatic environments.
Since chlorophyll-containing organisms are the first step in most food
chains, the health and /or abundance of these primary producers will
have cascading effects to all higher organisms. Therefore, the
determination of chlorophyll concentration is one of the key indices in
monitoring the health of any natural system.
Chlorophyll measurements are also used to directly
monitor phytoplankton populations. Examples include, but are not
limited to, the monitoring of chlorophyll in natural marine and
freshwater environments, reservoirs, water and sewage treatment plants,
and aquacultural systems.
Q: How do fluorometers detect and quantify chlorophyll a
in water?
A: Fluorescence is the phenomena of some compounds to absorb
specific wavelengths of light and almost instantaneously emit longer
wavelengths of light. Chlorophyll a naturally
absorbs blue light and emits, or fluoresces, red light. Fluorometers
detect chlorophyll a by transmitting an excitation
beam of light in the blue range (440nm for extracted analysis and 460nm
for in vivo analysis) and by detecting the light
fluoresced by cells or chlorophyll in a sample at 685nm (red).
Generally, this fluorescence is directly proportional to the
concentration of the material in question.
Q: What is the difference between in
vivo, in vitro , and extracted
chlorophyll analysis?
A: In vitro (meaning 'in glass' and
referring to 'in an artificial environment or outside the living
organism') chlorophyll analysis is another term for extracted analysis.
It entails the concentration of chlorophyll containing cells onto a
filter followed by the extraction of the chlorophyll a
from the cells. In vivo (meaning 'within a living
organism') chlorophyll analysis simply refers to the analysis of
chlorophyll in the natural environment or, in our case, in the living
algal cells.
Q: What is in vivo chlorophyll analysis?
A: In vivo chlorophyll analysis is the
fluorescent detection of chlorophyll a in living
algal and cyanobacterial cells in water. In this technique, the
excitation light from the fluorometer passes through the untreated
sample water and excites chlorophyll within the living cells of the
algae present. There are several factors that make in vivo
analysis a semi-quantitative measure at best. Environmental parameters,
physiology, morphology, light history and the presence of interfering
compounds all play a role in altering the relationship between
fluorescence and the concentrations of chlorophyll a.
Examples of interfering materials include other plant pigments,
degradation products, dissolved organic matter, and turbidity.
In vivo fluorescence data supplies information on the
relative distribution of chlorophyll concentrations and usually
correlate well with extracted chlorophyll a
samples.
In vivo detection has several very
useful applications. An example is the monitoring of general trends in
chlorophyll concentrations in real time. It is very easy to obtain
large amounts of data using in vivo instrumentation and is an excellent
means of following trends and estimating chlorophyll concentration.
With the introduction of the
SCUFA® submersible fluorometer,
verticals profiling and mooring applications are now possible. Other
examples of in vivo applications include continuous
monitoring along a ship's track using the 10-AU
configured with a flowcell and discrete sampling used
to monitor algal concentrations in natural or laboratory phytoplankton
populations. Examples of discrete sampling applications include
aquaculture and hatchery systems, water treatment facilities, reservoir
monitoring, and aquatic research (see the Aquafluor™).
If
water samples are taken, the in vivo data can be correlated to
extracted chlorophyll a data to estimate actual concentrations.
Otherwise, the in vivo data can be used as a relative measurement to
identify trends and patterns.
Q:
How do you calibrate a fluorometer for extracted chlorophyll a
analysis?
A:
Allow the fluorometer to warm-up for the time specified in the User's
Manual. Measure the fluorescence of each standard at sensitivity
settings that provide mid-scale readings (refer to your User's Manual
for proper calibration procedures). Follow directions under section
10.0 from E.P.A. Method 445.0 (Revision 1.2) for the calibration and
standardization procedure using the traditional acidification technique
or the non-acidification method. E.P.A. Method 445.0 calls for
filtering onto glass fiber filters (GFF) filters and grinding of the
filters. This step may not be necessary in some systems and tests
should be run to compare extractions with and without grinding.
Non-grinding techniques can use either GFF or membrane filters that
will dissolve in the solvent.
Q: What environmental factors interfere with in
vivo chlorophyll analysis?
A: Light, temperature, water quality, and dissolved
components can all have significant effects on fluorescent readings
independent of the chlorophyll concentration. However, all of these
factors can be controlled and/or corrected to a degree if the user is
aware of their effects.
Temperature
has an inverse relationship with fluorescence. For example, in a
vertical profile, as the temperature decreases, the fluorescence will
increase independent of chlorophyll concentration. The in vivo
chlorophyll fluorescence response changes at a rate of 1.4% per
°C. A temperature drop of 10 °C in a vertical profile
would result in a 14% overestimation of chlorophyll at the coldest
point. Turner Designs' field and submersible instruments have the
capability to automatically compensate for temperature effects.
Light
history can have significant affects on the fluorescence in algal
cells. For example, at low light levels, algal cells can optimize the
light uptake by pushing chloroplasts to the outer edge of the cell or
by producing more chlorophyll per cell. Both of these responses can
result in data increases the fluorescence signal while the algal
biomass may be unchanged. To lessen the effects, opaque hose should
always be used when sampling natural waters with a field fluorometer.
The transport time of the water in the hose will dark-adapt cells to an
extent, significantly reducing fluorescence error caused by variations
in the light history of the cells.
Dissolved
organic matter (DOM), chlorophyll degradation products (pheophytins),
chlorophyll b & c and
turbidity can also falsely increase the chlorophyll a
fluorescence signal. If the dissolved interfering compounds are
suspected to be significant, it is worth conducting a quick study to
look at the effects by comparing the fluorescence from filtered and
non-filtered water samples or from noting the fluorescence signal below
the photic zone and using this value as a blank.
Q: What is the effect of varying species on fluorometric
chlorophyll analysis?
A: Different species of phytoplankton have varied
morphologies (cell packaging), physiological states, size, and
chlorophyll a : carbon ratios. All of these factors
can affect the fluorescence emitted from a cell under a specific
excitation light intensity. Luckily, in natural environments, the
phytoplankton assemblage is diverse enough that much of the variation
in fluorescence resulting from the factors above balance eachother out
so the net effect on the fluorescence reading is reduced. Nevertheless,
the researcher needs to be aware of the potential for error in readings
caused by these factors to better interpret in vivo
data.
Q:
How does other chlorophylls and degradation products affect extracted
chlorophyll analysis?
A: All chlorophyll pigments and their degradation products
(pheophytins) have their own unique excitation and emission spectra.
Unfortunately, these spectra overlap significantly with the
fluorescence spectra of chlorophyll a due to the
similar chemical structure (see Figure 2). Due to the overlap in
fluorescence spectra, the presence of one can result in an interference
in the measurement of another, resulting in an under or overestimation
of the pigment in question.
In
the case of chlorophyll a, interfering pigments
to be
aware of are chlorophyll b, chlorophyll c, and
pheophytina(see figure 3). Chlorophyll b causes
the most interference in freshwater systems with high concentrations of
chlorophytes and/or prasinophytes and in marine systems with high
concentrations of prochlorophytes. In chlorophyll extraction, the
interference results during the acidification step of the traditional
extraction technique. Chlorophyll b undergoes a
wavelength shift when acidified, resulting in an underestimation of
chlorophylla and an overestimation of pheophytin.
In environments with high chlorophyll b
concentrations, we strongly recommend using the Welschmeyer
(non-acidification) method.
High
concentrations of chlorophyll c can result in a
slight overestimation of chlorophyll a and an
underestimation of pheophytin a sometimes even resulting in negative
pheophytin readings. It has been reported that a chla
:chlb ratio of 1 : 1 , which is the highest ratio
which could occur in nature, would result in achla overestimation
of 10%.
High
concentrations of any of the interfering compounds will result in an
increase to in vivo chlorophyll readings. The
magnitude of the increase will depend on the instrumentation used.
Instruments using filters with a wider bandpass will be more effected
by interfering compounds than instrument with narrower bandpass
filters.
Q:
What is a secondary standard?
A: A secondary standard is used as an alternative to a
primary calibration standard. It is often used when primary standards
are expensive, difficult to obtain, or unstable. In the case of
chlorophyll a, a secondary standard can be used the majority of time
for calibration because liquid chlorophyll a standards are expensive,
time-consuming, and photosensitive. To properly use a secondary
standard, you must first calibrate with a primary liquid standard of
the fluorophore of interest. You can then obtain the equivalent value
of the secondary standard. Henceforth, you may calibrate using the
secondary standard using the value you obtained for it initially. An
occasional calibration using a primary standard to recheck the
stability is recommended.
In
the past, secondary standards have been more stable than the primary
standards they mimic, but have still required special storage and
handling conditions with relatively short lifetimes. Examples of these
include coproporphrin and fluorescent dyes such as Rhodamine WT.
Turner
Designs has developed a solid secondary standard that is stable under
ambient light and temperatures with no special treatment or storage
required. This new secondary standard will greatly reduce time, cost,
and trouble in fluorescent chlorophyll analysis procedure.
Q:
What is the best solvent and procedure for the extraction of
chlorophyll a?
A: The
most commonly used extraction solvent is a 90% acetone 10% DI water
solution. Other solvents, such as methanol, ethanol and acetone/DMSO
mixtures are also commonly used and can improve extraction efficiency
with specific phytoplankton or may be found useful for the extraction
of sediment samples.
There
is no 'best' solvent or procedure for chlorophyll extraction. Several
work well and have their own pros and cons. The E.P.A. Method 445.0
describes the recommended step-by-step process for analysis using 90%
acetone. There are many factors in the extraction process that can lead
to different results. Several examples of these factors will be
discussed below in hopes reducing some of the variability.
An
excellent resource for sample collection, storage and extraction
methods is the UNESCO publication, Phytoplankton Pigments in
Oceanography.
Water
collection, storage, and filtration:
Because pigment is being extracted from living cells, it is critical to
use consistent techniques of obtaining water, filtering and storing
filters. The living cell is sensitive to changes in the environment
such as temperature and light. Conditions leading to cell death or
damage will affect chlorophyll concentrations.
In
the collection of water samples, it is important to make certain that
the collection containers are clean of all chemicals. They should be
rinsed several times in the sample water. Once collected, if samples
cannot be filtered immediately, they should be stored quickly on ice in
the dark. The time between collection and filtration should be as brief
as possible and should not exceed 4 hrs.
Specifics
on the recommended filtration and storage procedure can be found in
E.P.A. Method 445.0. An excellent resource for sample collection,
storage and extraction methods is the UNESCO publication, Phytoplankton
Pigments in Oceanography.
Q: How do I take and store discrete water samples in the
field?
A: For discrete in vivo analysis, water
samples should be measured as soon as possible after collection. The
same time constraints should be placed on the filtering of water
samples that are to be used for extracted analysis. From the time of
collection to measurement, the samples should be stored in the dark on
ice. Remember that the cells are living and significant time in a
container will alter the physiological state of the algal cells,
resulting in misrepresented chlorophyll data compared to the natural
situation.
Discrete
samples need to be kept at the same temperature. This is most easily
accomplished through the use of a water bath. The bath should be
covered from direct light. When the samples are being measured in the
fluorometer, a 'time in the instrument' must be established. Use the
discrete sample averaging function on the 10-AU or TD-700 Fluorometer
or wait for the reading to stabilize (~10 seconds) and record the
fluorescence. If this time is not monitored the heat and light in the
instrument will cause fluorescence to change.
When
developing you own sample collection and storage procedures, it is
recommended to run your own experiment to test change in fluorescence
over a given transport time by analyzing samples from a given sample at
hour or half-hour intervals.
Q: How do I
calculate actual chlorophyll a concentrations from
my fluorometric data?
A: ACIDIFICATION METHOD
Prior
to running sample on the fluorometer, the instrument must be calibrated
with a pure chlorophyll a standard and the maximum
acid ratio must be determined by measuring the fluorescence of the
standard before and after acidification. If afluorometer other than a
digital Turner Designs instrument is being used, the fluorometer
sensitivity coefficient may also need to be determined. For further
information please refer to EPA Method 445.0.
chl
a =K (Fm/ Fm -1) x (Fb-Fa)
x (v/V)
pheo a =K (Fm/ Fm -1) x
[(Fm x Fa - Fb)]
x (v/V)
If
necessary, the result can be multiplied by a dilution factor.
where:
K= sensitivity coefficient, equal to 1 on 10-AU
Fm = max acid ratio Fb/Fa of pure chlorophyll a standard
Fb = fluorescence before acidification
Fa = fluorescence after acidification
Fo = fluorescence signal of sample
v = extract volume (L)
V= volume filtered (L)
NON-ACIDIFICATION
METHOD (Chlorophylla concentration only)
1)
Collect fluorescence data (one number/sample)
** DO NOT ACIDIFY**
2) Plug data into following equation:
chl a = (Fo x
v)/ V
where:
Fo = fluorescence signal of sample
v = extract volume (L)
V= volume filtered (L)
Q: Does the
E.P.A. approve fluorometric chlorophyll analysis?
A: Yes, the
E.P.A. has published Method 445.0 which covers the
in vitro(extraction) fluorometric analysis of chlorophyll a.
In the most recent revision (Rev 1.2 Sept., 1997), the E.P.A. also
approves the use of the non
-acidification method, which is less susceptible to interfering
compounds such as chlorophyll b. This filter kit
supplies only chlorophyll a concentrations with no
information on pheophytin concentration.
Q: How can I
compare chlorophyll data obtained through different measurement
techniques?
A:
All detection instrumentation used in chlorophyll analysis will result
in chlorophyll concentrations that are directly comparable. A
side-by-side comparison between a fluorometer and a spectrophotometer
is easily done but would require dilution of the chlorophyll sample to
put it within the linear range of the fluorometer. A sample that is in
range on a spectrophotometer will be over-range on a fluorometer.
Q: Why use a
fluorometer over a spectrophotometer for extracted chlorophyll
analysis?
A: Benefits of
fluorescence over spectrophotometry include the capability of
in vivo detection, sensitivity, durability, versatility
(accepts a wide range of discrete sample cells and flow cells, accepts
AC or DC power, and the user can choose and quickly change between many
optical kits), ease of use, stability, ease of transport, and a small
footprint.
For
oceanographic research, the greater sensitivity of fluorescence results
in less time and work in the analysis because much less water must be
filtered for extracted analysis. The superior sensitivity also enables
in vivo detection of chlorophyll concentration of
<1µg/L.
Freshwater
researchers now have an extremely accurate and easy way to measure
chlorophyll a even with high chlorophyll b
concentrations using the non-acidification optical kit. Fluorometers
also allow for in-line monitoring to collect data in real time.
Q: What are
the chlorophyll detection limits of Turner Designs Fluorometers?
A: The TD-700
Laboratory Fluorometer and the 10-AU Field Fluorometer with a red
sensitive photomultiplier tube (PMT), have extracted chlorophyll
detection limits of 0.02µg/L using a 13mm diameter test tube
and 0.01µg/L using a 25mm test tube. The SCUFA®
Submersible Fluorometer can detect in vivo
chlorophyll concentrations to 0.02µg/L.
|